11 technical data, 1 power supply, 2 ambient conditions – Eppendorf BioSpectrometer kinetic User Manual
Page 85: 3 weight/dimensions, Power supply, Ambient conditions, Weight/dimensions

85
Technical data
Eppendorf BioSpectrometer
®
kinetic
English (EN)
11
Technical data
11.1
Power supply
11.2
Ambient conditions
Do not expose to direct sunlight.
11.3
Weight/dimensions
Power supply
100 V to 240 V ±10 %, 50 Hz to 60 Hz
Overvoltage category
II
Degree of pollution
2
Power consumption
Maximum power consumption according to name plate: 50 W
Approx. 30 W during operation
Approx. 5 W with the display dimmed and temperature control
switched off
Permitted mains interruption
Approx. 10 ms at 90 V
Approx. 20 ms at 230 V
Protection class
I
Fuses
T 2.5 A/250 V, 5 mm × 20 mm (2 pcs.)
Operation
Ambient temperature: 15°C to 35°C
Rel. humidity: 25% to 70%
Air pressure: 86 kPa to 106 kPa
Air pressure
Use up to an altitude of 2000 m above MSL
Weight
5.3 kg
Dimensions
Width: 295 mm
Depth: 400 mm
Height:150 mm
Space required
Width: 500 mm (with thermal printer: 750 mm)
Depth: 500 mm