Eppendorf BioSpectrometer kinetic User Manual
Page 46

Methods
Eppendorf BioSpectrometer
®
kinetic
English (EN)
46
Multi λ
Results display:
• Absorbance values at the wavelengths
Additional data ([Data] softkey):
• Only for dilution or with cuvettes other
than 10 mm: absorbance value before
the conversion.
Scan
Results display:
• Scan (graph with absorbance
wavelength display)
• Navigate between the measuring
points in the graph with and .
Routine
main group
Nucleic acids
Results display:
• Concentration result with absorbance
at the measuring wavelength
• If activated in the parameters: Ratios
A260/A280
• If activated in the parameters: Scan.
Navigate between the measuring
points on the graph which are used for
the result calculation with and .
Additional data ([Data] softkey):
If the corresponding parameters have
been activated:
• Absorbance value for 280 nm.
• Ratios A260/A230 and absorbance
value for 230 nm.
• Absorbance value for the background
wavelength.
Method group Results display
Explanation