2 variable regulators, 3 mode of operation – Eppendorf FemtoJet User Manual
Page 15

13
3 Mode of operation
Slight deviations between the nominal value (specified value) and
displayed compensation pressure p
c
(actual value) may occur for
technical reasons. The relevant value is the actual value measured.
3.1.2 Variable regulators
The variable regulators are set so that when they are turned slowly, the
values are changed by 1 hPa / 0.1 second per notch (0.01 – 0.02 psi).
If the regulators are turned faster, the values are altered in larger
increments.
The following settings are used as a
basis for initial injection experiments
using Femtotip
/ Femtotip
II:
Display:
p
i
left-hand variable regulator: 150 hPa
t
i
central variable regulator:
0.5 s
p
c
right-hand variable regulator: 50 hPa
Variable regulators:
AUTO
n= 0
pc= 50 hPa
pi[hPa]
ti[s]
pc[hPa]
150
0.5
50