6 technical data – Eppendorf FemtoJet User Manual
Page 42

40
Voltage/Frequency:
100 V – 240 V ± 10 %, 50 – 60 Hz
Set voltage:
adapts automatically
Fuse for 100 – 240 V:
T1.0 A, 250 V (2 pcs.)
Power consumption:
50 W
Max. current consumption: < 0.5 A
Protection class:
I
Overvoltage category:
II
(IEC 61010-1)
Pollution degree:
2 (IEC 664)
Ambient temperature:
15 °C – 35 °C
Ambient relative humidity:
max. 70 %
Weight: 6.2
kg
Dimensions:
Width:
220 mm
Depth:
170 mm
Height: 280
mm
Injection time t
i
:
0.0 – 99.9 s
in 0.1 s increments
Injection pressure p
i
:
0; 5 – 6,000 hPa (87 psi),
in 1 hPa (0.01 – 0.02 psi) increments
Compensation pressure p
c
: 0; 5 – 6,000 hPa (87 psi),
in 1 hPa (0.01 – 0.02 psi) increments
Regulating accuracy:
1 % of value set in each case,
at least ± 1 hPa
Clean rinsing pressure:
max. approx. 7,000 hPa (102 psi)
RS 232 computer
Baud rate: 9,600
interface:
Start bit:
1
Data bits:
8
Parity: non
Stop bit:
2
The device is approved with the
sign.
U.S. Pat.No. 4,836,244
Technical specifications subject to change!
6 Technical data
06_TechnDaten_en.fm Seite 40 Mittwoch, 27. September 2006 5:14 17