Appendix a – Eppendorf FemtoJet User Manual
Page 52

50
Appendix A
If the command is faulty, an error message is returned instead of the
answer:
"ERR=eee"
ERR:
Error
=: Separation
eee:
Error number
Lower and upper case letters can be used.
Command to enter
parameters
[C]
Code
[xxx]
Input
parameter
[ssssssssss]
Meaning of
input parameters
Short acoustic signal
003
1 – 254
Number of signals
Long acoustic signal
004
1 – 254
Number of signals
Automatic injection
(inject at t
i
)
010
0; 1 – 999
Injection time
in 1/10 seconds
0: injection time set on
the FemtoJet
Set injection counter n
to 0
012
– – –
No parameter input
Perform Clean
013
1 – 100
Time in 1/10 seconds
Execute Standby
014
0 or 1
0 = short
1 = long
Switch acoustic signal
on/off
040
0 or 1
Status:
0 = off / 1 = on
Set pressure unit
041
0 or 1
Unit:
0 = hPa / 1 = psi
Set injection time t
i
042
1 – 999
Injection time
in 1/10 seconds
A_Bedien_Computer_en.fm Seite 50 Mittwoch, 27. September 2006 5:26 17