5 troubleshooting – Eppendorf FemtoJet User Manual
Page 39

37
5 Troubleshooting
Error 25
No drop in working pressure
Possible cause: Technical defect
Solution: Service
Error 26
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.
Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.
Error 27
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.
Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.
ERROR 025
Working pressure
does not fall
INJECT = Continue
ERROR 026
Working pressure
leak is too high
INJECT = Continue
ERROR 027
Working pressure
leak is too high
INJECT = Continue