Other error numbers as well as the error cause and solution are described
in Section 5, "Troubleshooting". During remote control, no error texts
appear in the FemtoJet
display.
Error no.
during
execution ERR
= [eee]
Cause
Solution
070
FemtoJet
is not in routine
mode.
Check FemtoJet
.
Bring FemtoJet
into
routine operation.
Eliminate error.
071
FemtoJet
is not ready.
Check FemtoJet
or
micromanipulator.
As necessary, wait or
eliminate error.
072
The command is not finished
within the scheduled time.
Check FemtoJet
or
micromanipulator.
Eliminate error.
073
This command cannot be
executed at the moment,
because of insufficient storage
pressure.
See: Warning 001
Check FemtoJet
.
Eliminate error.
074
The connected micro-
manipulator did not respond.
See: Warning 003
Check micro-
manipulator.
Eliminate error.