5 Maintenance

16

Eppendorf Research® plus — Operating manual

5.2.2

Disinfection



Carefully wipe off the outer surfaces with disinfectant, DNA/RNA decontamination agents or

70% isopropanol.



You can wipe the outside of the tip cone and the ejector sleeve with a sodium hypochlorite

solution of 4%.



After the exposure time of the sodium hypochlorite solution has elapsed, remove thoroughly

with demineralized water.

The Research plus pipettes may be temporarily subjected to the UV light of a sterile bench ( ≥254

nm).

5.3 Replacing O-rings

The 100 μL and 300 μL multi-channel lower parts are equipped with O-rings. These are wear

parts. Replace old, worn or damaged O-rings. Defective O-rings result in the incorrect positioning

of the tips and in dispensing errors.

5.3.1

Removing the O-ring

1. Push the opening of the O-ring tool (included in the delivery package) against the tip cone

from the side so that the sharp edge in the opening of the O-ring tool is positioned on top of

the o-ring. Do not slide the O-ring tool onto the cone!

2. Supporting the O-ring tool with your thumb, push it firmly against the tip cone.

This O-ring is cut in one place.

3. Remove the O-ring tool and the O-ring from the tip cone.

5.3.2

Mounting a new O-ring

1. Fit the mounting aid (shortened pipette tip, included in the delivery package of the O-rings)

on the tip cone.

2. Push the new O-ring over the tip onto the tip cone.
3. Check that the pipette tips are correctly positioned. Ensure that the tips are tightly sealed and

properly aligned.

5