6 technical data, Technical data – Eppendorf Research plus User Manual

Page 36

Advertising
background image

6 Technical data

36

Eppendorf Research® plus — Operating manual

30 - 300 μL
Increment:
0.2 μL

orange
20 - 300 μL
55 mm

30 μL

± 2.5

± 0.75

± 0.7

± 0.21

150 μL

± 1

± 1.5

± 0.3

± 0.45

300 μL

± 0.6

± 1.8

± 0.2

± 0.6

100 - 1 000 μL
Increment:
1 μL

blue
50 - 1 000 μL
71 mm

100 μL

± 3

± 3

± 0.6

± 0.6

500 μL

± 1

± 5

± 0.2

± 1

1 000 μL ± 0.6

± 6

± 0.2

± 2

0.5 - 5 mL
Increment:
0.005 mL

purple
0.1 - 5 mL
120 mm

0.5 mL

± 2.4

± 12

± 0.6

± 3

2.5 mL

± 1.2

± 30

± 0.25

± 6

5.0 mL

± 0.6

± 30

± 0.15

± 8

1 - 10 mL
Increment:
0.01 mL

turquoise
1 – 10 ml
165 mm

1.0 mL

± 3

± 30

± 0.6

± 6

5.0 mL

± 0.8

± 40

± 0.2

± 10

10.0 mL

± 0.6

± 60

± 0.15

± 15

Research plus variable single-channel
Model

Test tip
epT.I.P.S. Color

code
Volume range
Length

Testing

volume

Error limits

Error

Systematic

Random

± %

± μL

± %

± μL

6

Technical data

Advertising