5 Maintenance

33

Eppendorf Research® plus — Operating manual

Error limits in accordance with ISO 8655-2 - Research plus variable multi-channel
Model

Test tip
epT.I.P.S. Color

code
Volume range
Length

Testing

volume

Error limits

Error

Systematic

Random

± %

± μL

± %

± μL

0.5 - 10 μL
Increment:
0.01 μL

medium gray
0.1 - 20 μL
40 mm

0.5 μL

± 48.0

± 0.24

± 32.0

± 0.16

1 μL

± 24.0

± 0.24

± 16.0

± 0.16

5 μL

± 4.8

± 0.24

± 3.2

± 0.16

10 μL

± 2.4

± 0.24

± 1.6

± 0.16

10 - 100 μL
Increment:
0.01 μL

yellow
2 - 200 μL
53 mm

10 μL

± 16.0

± 1.6

± 6.0

± 0.6

50 μL

± 3.2

± 1.6

± 1.2

± 0.6

100 μL

± 1.6

± 1.6

± 0.6

± 0.6

30 - 300 μL
Increment:
0.02 μL

orange
20 - 300 μL
55 mm

30 μL

± 26.0

± 7.8

± 10.0

± 3.0

150 μL

± 5.2

± 7.8

± 2.0

± 3.0

300 μL

± 2.6

± 7.8

± 1.0

± 3.0

5