Technical data, 1 research plus single-channel variable, 6 technical data – Eppendorf Research plus User Manual

Page 35

Advertising
background image

6 Technical data

35

Eppendorf Research® plus — Operating manual

6

Technical data

6.1 Research plus single-channel variable

Research plus variable single-channel
Model

Test tip
epT.I.P.S. Color

code
Volume range
Length

Testing

volume

Error limits

Error

Systematic

Random

± %

± μL

± %

± μL

0.1 - 2.5 μL
Increment:
0.002 μL

dark gray
0.1 - 10 μL
34 mm

0.1 μL

± 48

± 0.048 ± 12

± 0.012

0.25 μL

± 12

± 0.03

± 6

± 0.015

1.25 μL

± 2.5

± 0.031 ± 1.5

± 0.019

2.5 μL

± 1.4

± 0.035 ± 0.7

± 0.018

0.5 - 10 μL
Increment:
0.01 μL

medium gray
0.1 - 20 μL
40 mm

0.5 μL

± 8

± 0.04

± 5

± 0.025

1 μL

± 2.5

± 0.025 ± 1.8

± 0.018

5 μL

± 1.5

± 0.075 ± 0.8

± 0.04

10 μL

± 1.0

± 0.1

± 0.4

± 0.04

2 - 20 μL
Increment:
0.02 μL

light gray
0.5 - 20 μL L
46 mm

2 μL

± 5

± 0.1

± 1.5

± 0.03

10 μL

± 1.2

± 0.12

± 0.6

± 0.06

20 μL

± 1.0

± 0.2

± 0.3

± 0.06

2 - 20 μL
Increment:
0.02 μL

yellow
2 - 200 μL
53 mm

2 μL

± 5

± 0.1

± 1.5

± 0.03

10 μL

± 1.2

± 0.12

± 0.6

± 0.06

20 μL

± 1.0

± 0.2

± 0.3

± 0.06

10 - 100 μL
Increment:
0.1 μL

yellow
2 - 200 μL
53 mm

10 μL

± 3

± 0.3

± 1

± 0.1

50 μL

± 1

± 0.5

± 0.3

± 0.15

100 μL

± 0.8

± 0.8

± 0.2

± 0.2

20 - 200 μL
Increment:
0.2 μL

yellow
2 - 200 μL
53 mm

20 μL

± 2.5

± 0.5

± 0.7

± 0.14

100 μL

± 1

± 1

± 0.3

± 0.3

200 μL

± 0.6

± 1.2

± 0.2

± 0.4

6

Technical data