5 Maintenance

17

Eppendorf Research® plus — Operating manual

5.4 Research plus disassembly and assembly

5.4.1

Single-channel up to 1 000 μL

Removing the lower part

1. Keep the ejector pressed and remove the ejector sleeve .
2.

and : on the lower part, slide the ring marked "PUSH UP TO RELEASE" up by about

5 mm until the lower part comes off.

3.

: take the lower part out of the upper part.

Deactivating the spring-loading action: installing the locking ring
For the following dispensing tasks it can be helpful to deactivate the spring loading action of the

tip cone.

A pipette tip is to be used for a long period.

The pipette tip is slightly bended during dispensing.

Deactivating this action can also be helpful for pipette tips which are not designed for the

Eppendorf pipette tip cone.

The spring loading action of the tip cone can be deactivated by installing

the locking ring. The locking ring is included in the delivery package.

1. Push the black locking ring onto the lower part from the top. Slightly

squeeze the clamps on the lower part when doing so.

2. Insert the lower part into the upper part until it engages audibly.
3. Keep the ejector pressed. The ejector rod protrudes from the upper

part.

4. Fit the ejector sleeve onto the ejector rod. It is fitted correctly if it

engages slightly.

Remove the locking ring to reactivate the spring-loading action.

PUSH UP TO RELEASE

1

2

3

4

1

2

3

4

5