Maintenance
Combitips advanced

®

English (EN)

16

5

Maintenance

5.1

Cleaning

5.1.1

Autoclaving the Adapter advanced

Rinse the Adapter advanced with water.

Autoclave the Adapter advanced for 20

min at an overpressure of 1

bar and 121

°C. The Adapter

advanced can be autoclaved up to 100 times.

5.1.2

Autoclaving the Combitips advanced Rack

Autoclave the empty Combitips advanced Rack for 20

min at an overpressure of 1

bar and 121

°C.

Empty Combitips advanced Rack can be autoclaved up to 100 times.

6

Technical data

6.1

Ambient conditions

Combitips advanced are not autoclavable. Adapter advanced and Combitips advanced Rack
are autoclavable.

Ambience

For indoor use only.

Ambient temperature

4

°C

40

°C

Relative humidity

10

%

95

%,

non-condensing

Atmospheric pressure

79.5

kPa

166

kPa