Technical data

Combitips advanced

®

English (EN)

19

6.3

Errors with the Multipette plus/Repeater plus

Test conditions and test evaluation in compliance with ISO

8655,

Part

6. Analytical balance with

evaporation protection inspected by the office of weights and measures.

• Number of determinations: 10

• Use of water in accordance with ISO

3696

• Inspection at 20

°C – 25

°C

±0.5

°C

• Dispensing on the inner wall of the tube

Combitip advanced

Volume range

Error limits

Error

Systematic error

Random error

±

%

±

µL

±

%

±

µL

0.1

mL
white

2

μL

±1.6

±0.032

±3.0

±0.06

20

μL

±1.0 ±0.2 ±2.0

±0.4

0.2

mL
light blue

4

μL

±1.3

±0.052

±2.0

±0.08

40

μL

±0.8

±0.32

±1.5

±0.6

0.5

mL
purple

10

μL

±0.9

±0.09

±1.5

±0.15

100

μL

±0.8 ±0.8 ±0.6

±0.6

1

mL
yellow

20

μL

±0.9

±0.18

±0.9

±0.18

200

μL

±0.6 ±1.2 ±0.4

±0.8

2.5

mL
green

50

μL

±0.8 ±0.4 ±0.8

±0.4

500

μL

±0.5 ±2.5 ±0.3

±1.5

5

mL
blue

100

μL

±0.6 ±0.6 ±0.6

±0.6

1

000

μL

±0.5

±5.0

±0.25

±2.5

10

mL
orange

200

μL

±0.5

±1.0

±0.6

±1.2

2

000

μL

±0.5

±10

±0.25

±5.0

25

mL
red

500

μL

±0.4

±2.0

±0.6

±3.0

5

000

μL

±0.3

±15

±0.25

±12.5

50

mL
light gray

1

000

μL

±0.3

±3.0

±0.5

±5.0

10

000

μL

±0.3

±30

±0.3

±30

The test volumes for the systematic and random errors of the Multipette (X)stream/Repeater
(X)stream comply with the requirements of ISO 8655, part 5.