2 errors with the multipette m4/repeater m4, Errors with the multipette m4/repeater m4 – Eppendorf Combitips advanced User Manual

Page 17

Advertising
background image

Technical data

Combitips advanced

®

English (EN)

17

6.2

Errors with the Multipette M4/Repeater M4

Combitip advanced

Testing volume

Error limits

Error

Systematic error

Random error

±

%

±

µL

±

%

±

µL

0.1

mL
white
Increment:
1

μL

2

μL

±1.6

±0.032

±3.0

±0.06

20

μL

±1.0 ±0.2 ±2.0

±0.4

0.2

mL
light blue
Increment:
2

μL

4

μL

±1.3

±0.052

±2.0

±0.08

40

μL

±0.8

±0.32

±1.5

±0.6

0.5

mL
purple
Increment:
5

μL

10 μL

±0.9

±0.09

±1.5

±0.15

100

μL

±0.8 ±0.8 ±0.6

±0.6

1

mL
yellow
Increment:
10 μL

20

μL

±0.9

±0.18

±0.9

±0.18

200

μL

±0.6 ±1.2 ±0.4

±0.8

2.5

mL
green
Increment:
25

μL

50

μL

±0.8 ±0.4 ±0.8

±0.4

500

μL

±0.5 ±2.5 ±0.3

±1.5

5

mL
blue
Increment:
50

μL

100

μL ±0.6 ±0.6 ±0.6

±0.6

1

000

μL

±0.5

±5.0

±0.25

±2.5

10

mL
orange
Increment:
0.1

mL

200

μL
0.2

mL

±0.5 ±1.0

±0.6

±1.2

2

000

μL

2

mL

±0.5 ±10

±0.25

±5.0

25

mL
red
Increment:
0.25

mL

500

μL
0.5

mL

±0.4 ±2.0

±0.6

±3.0

5

000

μL

5

mL

±0.3 ±15

±0.25

±12.5

50

mL
light gray
Increment:
0.5

mL

1

000

μL

1

mL

±0.3 ±3.0

±0.5

±5.0

10

000

μL

10

mL

±0.3 ±30

±0.3

±30