FUSE REPLACEMENT

1. Press the ON(1)/OFF(0) switch to the OFF(0) position.

2. Remove the AC line cord from the AC power receptacle.

3. Remove the AC line cord from the power entry module at the rear of the illuminator. The fuse drawer is part

of the power entry module. The drawer is located directly beneath where the AC line cord plugs in.

4. Pull out the fuse drawer. Remove the blown fuse that is positioned closest to the illuminator and discard. The

second fuse is the spare.

5. Place the replacement fuse into the fuse drawer. The fuse will work in either orientation.

6. Push the fuse drawer until it "clicks" into position.

7. Attach the AC line cord to power entry module at the rear of the illuminator. The illuminator is now ready for

service.

WARNING

Replace the fuse with the correctly

rated fuse as listed on the label

on the back of the illuminator. Use of

an improper fuse can create

a hazardous situation.

®

Fiber-Lite

LAMP REPLACEMENT
1. Follow steps 1-4 of Lamp Module Replacement.
2. Check the lamp assembly to verify that the lamp and socket are cool before proceeding.
3. Lift and remove the lamp from the lamp holder by grasping the rear of the lamp adjacent to the lamp socket.
4. Remove the lamp from the socket by holding the lamp socket and gently pulling the lamp reflector.
5. Discard the lamp.
6. Insert the replacement lamp into the lamp socket. CAUTION: Do not touch the interior of the lamp reflector, the

lamp envelope or the lamp pins with your fingers. Touching the interior of the lamp reflector, the lamp envelope
or the lamp pins will result in significant shortening of the lamp life. Handle the lamp only by the exterior of the
reflector or the area adjacent to the pins.

7. Insert the replacement lamp and socket into the lamp holder.

8. Follow steps 7-9 of Lamp Module Replacement.

WARNING

Risk of electrical shock.

Remove power plug before lamp

replacement and wait for

hot lamp to cool.