Vacuum supply, Electric branching, Set-up – Heidolph Vacuum Switchbox User Manual

Page 46: Install cooler holder and condensate cooler, Rotavac vario control

Advertising
background image

46

E

1.3. Vacuum

supply

- Connect vacuum with stud (12)


1.4. Electric

branching

- Before connecting with your local power supply make sure that:

mains voltage is identical with data plate (refer to "safety Information"; refer to "General" if
plug and outlet do not fit together.

- Connect power cord with outlet.

2. ROTAVAC vario control

Caution: ROTAVAC vario control may be used together with the LABOROTA 4002/4003
control vario evaporator only !


2.1. Set-up

- Locate items on solid and plain surface. Item must be horizontal.

2.2. Install cooler holder and condensate cooler

- If your ROTAVAC vario control is used together

with the ROTACOOL reflux cooler, you may directly
place your ROTAVAC vario control on top of the
ROTACOOL.

- Using your ROTAVAC vario control without

ROTACOOL, an extra foot (15) has to be installed
for the condensate cooler.

- Refer to picture and install extra foot (15) with 4 ea.

screws (16), using hex pin from
you hardware kit.

- Remove fairing (17) by

unthreading two screws

- Install bracket (1) for condensate

cooler (2) on the rear of the
vacuum pump in a slot provided
to this end. Introduce 2 ea.
square nuts in slot.

- Re-install fairing (17).
- Remove coupling nut (10) from

condensate cooler (Attention,
retainer underneath coupling
nut!)

- Locate condensate cooler (2) in

holder

- Secure aftercooler with rubber

band (3).

- Install receiving flask (4) and

secure with clamp (5).

- Fit red coupling ring (10) and

seal (9) on right-hand end of
PTFE hose (7). Plug this line

ROTACOOL

14

3

2

1

17

12

6

11

4

5

10

15

16

7

18

Advertising