Heidolph Vacuum Switchbox User Manual

Page 66

Advertising
background image

66

E

Clean and inspect pump heads

Only ROTAVAC vario control:
Use box-end spanner (SW 17) and loosen coupling ring of hose beside gas-ballast valve.
Remove hose fitting from hose with ¼ revolution with box-end spanner (SW 15). Do not remove
fittings from pump head.
¾

Leaks may be found on re-assembly.




ROTAVAC vario control, ROTAVAC vario tec und ROTAVAC valve tec:
Loosen 6 ea. straight hex-recessed screws on pump head and lift off cover ass'y. (cover, internal
part, and lid):
¾

Never attempt to loosen parts with sharp tools

(screwdriver), use rubber hammer or compressed air
instead.







To inspect valves, carefully remove cover from lid. Denote
location of valves and remove.
¾

Replace damaged valves.

Remove contamination with cleaning agent.
Inspect diaphragm for damage and replace on condition.

Explosion view: pump head

housing

bearing

diaphragm
back-up disc

diaphragm

diaphragm
cocking disc
with screw

cover

piston rod

valve

lid

Advertising