Inspect screwed fitting of hoses on pump heads.

DISASSEMBLE, FORWARD & STORE


Disassemble

Turn item OFF and disconnect mains plug.

Transportation and store

1. Store item and its components in the original box to avoid damage in transit. Close box with

adhesive tape.

2. Store item in a dry place.

Caution: avoid shocks an vibration in transit.

DISPOSAL


For disposal, please comply with your local or national regulations.
Split by metal, plastic, etc.
Packing material to be treated as described above (material split).

TROUBLESHOOTING

Vacuum pumps

defect found

reason

correction

pump won't run or
starts running and
stops right away

¾

electricity missing, voltage too

low ?


¾

ROTAVAC vario control and

ROTAVAC vario tec:
triggering signal for
motorspeed missing?

¾

overpressure in exhaust line?

¾

motor

overload?

¾

check voltage, connect

electricity

¾

check triggering signal




¾

open exhaust line.

¾

allow motor to chill, check for

reason, turn ON again

low suction power

¾

centering ring not seated

correctly

¾

long, thin line?

¾

condensate in pump?

¾

¾

deposits in pump ?

¾

diaphragm and valves broken?

¾

emission of gas, vapor in

process ?

¾

ROTAVAC vario control and

Rotavac vario tec:

¾

pump runs hot (decrease

speed)?

¾

check pump check fittings. line

and recipient

¾

increase cross section.

¾

run pump with intake stud

open for a minute

¾

clean and check pump heads

¾

replace diaphragm and/or

valves

¾

check process parameters



¾

improve cooling, or reduce

suction pressure