3.6.10 Description of the RS-3000 I/O:

Table 3-2

Back-panel 15-pin D-sub Male Pin Descriptions

PIN
NUMBER

DESCRIPTION of Connector Labeled T3000 Control

I/O
direction

1

When this line is high, the RS-3000 will indicate an Open Boat
failure

Input
w/ pull-up

2

Along with pin 10, this line provides the RS-3000 with +5 VDC
power

Input
power

3

Along with pin 11, this line provides the RS-3000 with digital
ground

Input
power

4

Monitors the voltage output from the T3000 power supply
( 1 VDC = 1.5 VRMS )

Input

5

Alternative to T3000 Reset button (falling edge, not level
sensitive)

Output
w/ pull-up

6

For future use, not implemented in software.

Input

7

A control voltage from the RS-3000 used to set the output
amperage of the T3000 (see note

1

)

Output

8

Not connected

N. A.

9

When this line is high, the RS-3000 will indicate an Over
Temperature failure

Input
w/ pull-up

10

Along with pin 2, this line provides the RS-3000 with +5 VDC
power

Input
power

11

Along with pin 3, this line provides the RS-3000 with digital
ground

Input
power

12

This line provides the RS-3000 with +12 VDC power

Input
power

13

A digital control voltage from the RS-3000 used to switch the
T3000 power supply output ON (w/ a digital low) or OFF (w/ a
digital high)

Output
w/ pull-up

14

This line tells the RS-3000 how to interpret the T3000 data: as
series wired or parallel wired. Digital high = series, digital low =
parallel.

Input
w/ pull-up

15

Monitors the amperage output of the T3000 ( 1 VDC = 100 Amps
RMS when parallel wired / 50 Amps RMS when series wired

Input


Note

1

: 10 VDC = 600 AMPs parallel wired [or 300 AMPs series] and has a linear relationship down to zero except for a

small value range between zero volts and a small fraction of a volt where AMPs will remain at zero.

Section 3

Page 3 - 5

Specifics