Rs-3000 – INFICON RS-3000 Power Supply Controller User Manual

Page 17

Advertising
background image

RS-3000

Power Supply Controller

instruments

Table 3-3

Back-panel 15-pin D-sub Female Pin Descriptions

PIN
NUMBER

DESCRIPTION of Connector Labeled External Control

I/O
Direction

1

Analog ground of the RS-3000 presented to remote controller
(used with Voltage control [pin #9])

ground
pass thru

2

Along with pin 10, this line presents the RS-3000 digital ground to
the remote controller

ground
pass thru

3

Interlock status indication: grounded (to pin 2) or digital low =
satisfied, open or digital high = interlock error (indicated by front
panel LED)

Input
w/ pull-up

4

Process abort (digital low = process abort)

Input
w/ pull-up

5

Process START event (digital falling edge = START event)

Input
w/ pull-up

6

Fault reset (digital falling edge = reset event)

Input
w/ pull-up

7

Remote out (digital high = remote enabled)

Output
w/ pull-up

8

Ready (digital high = ready enabled)

Output
w/ pull-up

9

Voltage control (remote voltage control, via RS-3000, of the
T3000)

Input

10

Along with pin 2, this line presents the RS-3000 digital ground to
the remote controller

ground
pass thru

11

Process timer reset (digital falling edge = reset event)

Input
w/ pull-up

12

Process STOP event (digital falling edge = STOP event)

Input
w/ pull-up

13

Remote/Local START mode

Input
w/ pull-up

14

Busy (digital high = active)

Output
w/ pull-up

15

Fault (digital high = active)

Output
w/ pull-up