1 specifications – INFICON Front Load Single Sensor User Manual

Page 17

Advertising
background image

1 - 7

PN

07

4-

15

6N

Front Load Single and Dual Sensors Operating Manual

1.5.1 Specifications

Maximum bakeout temperature

with no water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130°C

Maximum operating isothermal environment
temperature with minimum water flow . . . . . . . . . . . 400°C

Maximum sensor body
envelope without shutter . . . . . . . . . Standard Sensor

27.00 x 61.47 x 17.53 mm
(1.06 x 2.42 x 0.69 in.)
See

Figure 1-6 on page 1-10

.

Right Angle Sensor
28.19 x 26.92 x 26.92 mm
(1.11 x 1.06 x 1.06 in.)
See

Figure 1-8 on page 1-12

.

Water tube and
in-vacuum cable length . . . . . . . . . . Standard Length (E)

76.2 cm (30 in.) tubes,
3.175 mm (1/8 in.) OD seamless stainless
steel, 0.406 mm (0.016 in.) wall thickness.

Includes 78.1 cm (30.75 in.) in-vacuum
cable.

Extended Length (G)
121.9 cm (48 in.) tubes,
3.175 mm (1/8 in.) OD seamless stainless
steel, 0.406 mm (0.016 in.) wall thickness.

Includes 152.4 cm (60 in.) in-vacuum cable.

Crystal exchange . . . . . . . . . . . . . . . Front-loading, self-contained package for

ease of exchange

Mounting. . . . . . . . . . . . . . . . . . . . . . Two #4-40 tapped holes on the back of the

sensor body

Crystal size . . . . . . . . . . . . . . . . . . . . 14 mm (0.550 in.) diameter