INFICON Front Load Single Sensor User Manual

Page 83

Advertising
background image

5 - 7

PN

07

4-

15

6N

Front Load Single and Dual Sensors Operating Manual

5

At the feedthrough, measure resistance between center contact and shield of
the BNC connector (refer to

Figure 5-1

).

If resistance is less than 1 Ω, electrical continuity is good.

If resistance is more than 1 Ω, check the following before continuing to
step 6:

Verify that in-vacuum cable connections to sensor head and
feedthrough are tight. Do not overtighten.

Remove the crystal holder and examine the leaf spring inside the
sensor head cavity. The three prongs on the leaf spring should reach to
approximately 1 mm (0.039 in.) from the top of the sensor head cavity.
To check prong height, place a straight edge across the sensor head
cavity (as shown by

Figure 5-4

). If the prongs are not high enough,

gently bend each prong upward using a gloved finger or plastic
tweezers.

Figure 5-4 Leaf spring

6

Reinstall the crystal holder with ceramic retainer into the sensor.

7

Measure resistance between center contact and shield of the BNC.

If resistance is less than 1 Ω, electrical continuity is good.

If resistance is more than 1 Ω, continue to step 8.

8

Disconnect the in-vacuum cable from the sensor head and remove the crystal
holder.

Advertising
This manual is related to the following products: