INFICON Front Load Single Sensor User Manual

Page 85

Advertising
background image

5 - 9

PN

07

4-

15

6N

Front Load Single and Dual Sensors Operating Manual

10

Disconnect the in-vacuum cable from the feedthrough.

11

At the feedthrough, measure resistance between the center contacts of the
BNC and coaxial connectors (shown by

Figure 5-6

).

Figure 5-6 Feedthrough Resistance

If resistance is more than 0.3 Ω, contact INFICON (refer to

section 1.3, How

to Contact INFICON, on page 1-2

).

If resistance is less than 0.3 Ω, continue to step 12.

12

Replace the in-vacuum cable.

13

Reinstall the crystal holder with ceramic retainer and without the crystal into
the sensor.

14

Measure resistance between center contact and shield of BNC.

If resistance is less than 1 Ω, electrical continuity is good.

If resistance is more than 1 Ω, contact INFICON.

Advertising
This manual is related to the following products: