5 - 9
PN
07
4-
15
6N
Front Load Single and Dual Sensors Operating Manual
10
Disconnect the in-vacuum cable from the feedthrough.
11
At the feedthrough, measure resistance between the center contacts of the
BNC and coaxial connectors (shown by
).
Figure 5-6 Feedthrough Resistance
If resistance is more than 0.3 Ω, contact INFICON (refer to
to Contact INFICON, on page 1-2
If resistance is less than 0.3 Ω, continue to step 12.
12
Replace the in-vacuum cable.
13
Reinstall the crystal holder with ceramic retainer and without the crystal into
the sensor.
14
Measure resistance between center contact and shield of BNC.
If resistance is less than 1 Ω, electrical continuity is good.
If resistance is more than 1 Ω, contact INFICON.