Front Load Single and Dual Sensors Operating Manual

Water Quality . . . . . . . . . . . . . . . . . . Coolant should not contain chlorides as

stress corrosion cracking may occur.
Extremely dirty water may result in loss of
cooling capacity.

CAUTION

Do not allow water tubes to freeze. This may happen if the
tubes pass through a cryogenic shroud and the flow of
fluid is interrupted.

Air (SL-X1XXX only). . . . . . . . . . . . . 70 psi (gauge) {85 psi (absolute)}

(5.8 bar (absolute)) [584 kPa (absolute)]
(minimum)

80 psi (gauge) {95 psi (absolute)}
(6.5 bar (absolute)) [653 kPa (absolute)]
(maximum)

WARNING

Do not exceed 100 psi (gauge) {115 psi (absolute)}
(7.9 bar (absolute)) [791 kPa (absolute)].

Connection to excessive pressure may result in personal
injury or equipment damage.

1.5.4 Single Sensor Drawings

The following Single Sensor Outline and Assembly Drawings provide dimensions
and other relevant data necessary for planning equipment configurations.

Figure 1-6

. . . . . . Standard Crystal Sensor Outline

Figure 1-7

. . . . . . Standard Crystal Sensor Assembly (PN 750-211-GX)

Figure 1-8

. . . . . .Right Angle Crystal Sensor Outline

Figure 1-9

. . . . . .Right Angle Crystal Sensor Assembly (PN 750-213-GX)