Able, Ystem, Aults – INFICON MPI-16 Crucible Indexer User Manual
Page 44
MPI-16 & MPI-16H POCKET INDEXERS
4-10 Operation
8
Blinks – Firmware Fault
Description:
• The system determined that the internal program was not executing correctly.
Action:
• Reset the MPI-16. This fault is an indication of a firmware problem, and
should only occur if your unit has become damaged. If the fault is persistent,
contact INFICON for assistance.
9 Blinks – Brown-Out Fault
Description:
• The supply voltage has dropped to an unexpected level.
Action:
•
Reset the MPI-16. Check that your power supply is operating correctly. If
the fault is persistent, the problem could be due to the internal voltage
regulator of the MPI-16. If you are certain that your power supply is
operating correctly, contact INFICON for assistance.
Table 4-5 System Faults