4.4 Abort and Reset

The MPI-16 has a multi-purpose Abort/Reset switch on its main panel. An additional
switch providing the same functionality is located on the optional Remote Panel. This
button has a long debounce time (150 ms) to ensure that noise or accidental triggering
does not stop your process.

• Pressing the Abort/Reset button once will pause the MPI-16 and turn off the

Interlock.

• Pressing the Abort/Reset button again will resume operation from the point it

was interrupted.

• At any point, holding the button for two seconds will reset the MPI-16.


When the MPI-16 is paused with this button, the Fault LED will blink.