tira46e1 (2004-11) BPG402-SDv1.cp

Following is the Instance 2 with the subclass extension of the hot cathode ion
gauge part of the BPG402-SD. This instance is used to provide control and status
information for the hot cathode ion gauge part of the BPG402-SD.

Attribute

ID

Access

Rule

NV/

V

Name

DeviceNet
Data Type

Description of Attribute

Semantics of Values

3

Set/Cond.

→ below

NV

Data Type

USINT

Determines the Data Type
of Value and all related
attributes as specified in
this table.

→ "Semantics" section
[default] = INT

4

"Semantics"

NV

Data Units

ENGUNITS

Determines the Units con-
text of Value and all related
attributes.

→ "Semantics" section
[default] = Counts

5

Get

V

Reading
Valid

BOOL

Indicates that the Value
attribute contains a valid
value.

0 = invalid
1 = valid
(invalid: e.g., not warmed up yet)

6

Get

V

Value

INT or speci-
fied by Data
Type
if sup-
ported

Analog input value

The corrected, converted, calibrated
final value of the sensor.
→ "Semantics" section

7

Get

V

Status

BYTE

Alarm and Warning State
of this object instance

→ "Semantics" section

10

Get

NV

Full Scale

INT or speci-
fied by Data
Type
if sup-
ported

The Value of Full Scale for
the sensor.

The value of attribute Value corre-
sponding to the Full Scale calibrated
measurement of the sensor.

[default] = maximum allowable value
for the Data Type

25

Set

NV

Safe State

USINT

Specifies the behavior for
the Value for states other
than Execute

→ "Semantics" section
[default] = 0

26

Set

NV

Safe
Value

INT or speci-
fied by Data
Type
if sup-
ported

The Value to be used for
Safe State = Safe Value

→ "Semantics" section
[default] = 0

32

Get

NV

Overrange INT or speci-

fied by Data
Type
if sup-
ported

Specifies the highest valid
Value

The value above which attribute
Reading Valid is set to invalid.

[default] = maximum allowable value
for the Data Type

33

Get

NV

Under-
range

INT or speci-
fied by Data
Type
if sup-
ported

Specifies the lowest valid
Value

The value below which attribute
Reading Valid is set to invalid.

[default] = minimum allowable value
for the Data Type

88

Get

V

Degas
Status

BOOL

Indicates current degas
state

0

OFF

1

ON

89

Set

V

Active
Filament

BYTE

Indicates selected filament

Bit 0 =1

filament 1 selected

Bit 1 =1

filament 2 selected

91

Get

V

Emission
Current

REAL

Indicates setting level of
emission current in amps

92

Set

NV

Filament
Control

USINT

Defines the mode in which
the filament is used.

0

Auto Î The device switches the

filament from Fil1 to Fil2 and
from Fil 2 to Fil1 every time the
emission is switched on.

1

Manual Î The device uses the

filament which is selected by
the attribute ActiveFilament

93

Get

V

Emission
Status

BOOL

Indicates whether the
emission is turned ON or
OFF

0

OFF

1

ON

94

Get

V

Sensor
Warning

Structure of
Byte

default

0 → "Semantics"

95

Get

V

Sensor
Alarm

Structure of
Byte

default

0 → "Semantics"

2.4.2.3 Instance Attributes of

Instance 2 Hot Cathode
Ion Gauge

Advertising