61

h

Set Degas State Activates/deactivates degas mode according to

the parameter Degas State. Degas mode may
be terminated either automatically by device
timeout (3 min) or remotely by this service.


Parameter

Data Type

Semantics of Values

Degas

State

BOOL

0 switches Degas OFF
1 switches Degas ON

Degas is switched on only if the pressure is below 7.2E-6 mbar.

Service

Code

Service Name

Description of Service

62

hex

Set Emission

State

Turns the filament on and off according to the
parameter Emission State

Parameter

Data Type

Description

Semantics of Values

Emission

State

BOOL

→ below

0 turns emission OFF
1 turns emission ON

If the gauge is in the Emission User Mode “Manual”, please keep attention at the
following points:

The service SET EMISSION STATE

Î ON may be executed only, if the pressure

is below 2.4E-2 mbar. If the pressure is higher, an Object State Conflict error mes-
sage will be returned.

If the pressure rises above 3.2E-2 mbar, the gauge will switch off the emission
automatically. The gauge will then show the pirani value.

If the gauge is measuring in the ion gauge measuring range and the service Set
Emission State

Î OFF ist executed, the emission will be switched off and the mi-

nimum Pirani value 1E-3 will be displayed.