INFICON BPG402-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual

Page 6

Advertising
background image

6

tira46e1 (2004-11) BPG402-SDv1.cp

8

9

10

Position

Function

8

Potentiometer for Setpoint A threshold adjustment

9

Potentiometer for Setpoint B threshold adjustment

10

Fullscale adjustment push button

Device State

LED State

Description

Power Off

dark

No power applied to device.

Device Self-Test

flashing
green-Red

Device is in self-test.

Device Operational

green

Device is operating normally.

Unrecoverable
Fault

red

Device has detected an unrecoverable fault.
As stated previously, all module level faults
are considered as unrecoverable faults.

Minor Fault

flashing red

Device has detected a recoverable fault
(e.g. DeviceNet power supply not
connected).

The network status LED indicates the status of the DeviceNet communication link.

For this state

LED state

To indicate

Not Powered/Not Online

dark

Device is not online.
• The device has not completed the

Duplicate MAC_ID test yet.

• The device may not be powered,

look at Module Status LED.

On–line, Not Connected

flashing

green

Device is online but has no connections
in the established state.
• The device has passed the

Dup_MAC_ID test, is online, but
has no established connections to
other nodes.

• The device is not allocated to a

master.

Link OK Online, Connected

green

The device is online and has connec-
tions in the established state.
• The device is allocated to a Master.

Connection Time Out

flashing

red

One or more I/O Connections are in the
timed out state.

Critical Link Failure

red

Failed communication device. The
device has detected an error that has
rendered it incapable of communicating
on the network (Duplicate MAC ID, or
Bus–off).

1.4 Side View of the

BPG402-SD

1.5 Indicators and Switches

1.5.1 Module Status LED

1.5.2 Network Status LED

Advertising