D: typical start-up procedure – INFICON BPG402-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual

Page 30

Advertising
background image

30

tira46e1 (2004-11) BPG402-SDv1.cp

00 00 CA 42

h

(4-Byte, floating format)

42

h

CA

h

00

h

00

h

0 1 0 0 0 0 1 0

b

1 1 0 0 1 0 1 0

b

0 0 0 0 0 0 0 0

b

0 0 0 0 0 0 0 0

b

0

1000 0101

b

1 0 0 1 0 1 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0

b

Ш

Ш

Ш

-1

0

1+

⎟⎟

⎜⎜

23

b

2

0000

0000

0000

0000

1010

100

= 1+

4849664

8388608

d

d



Ш

Ш

Sign =

Exponent =

Mantissa =

1

133

1.578125

1

Ч

2

(133-127)

Ч

1.578125 = 101

The start up of a device is divided into the steps:
• Allocation process
• Setting of the EPR attribute
• Choice of the input and output assemblies

Send an allocation string as defined in the DeviceNet specification to the device
you want to allocate.

Set the bits in the allocation choice byte to 1 for these connections you want to use.

Sign

8-Bit exponent

23-Bit mantissa

Example

Number received

1. Convert sequence of the

HEX words

2. Separate into bytes

3. Calculate

Converted number

D:

Typical Start-Up
Procedure

Allocation process

Advertising