2 gas type dependence, 3 switching functions sp1, sp2 – INFICON PCG55x ATM to Medium Vacuum Gauge User Manual

Page 40

Advertising
background image

40

tina56e1-d (2014-01)

EEPROM error

Sensor error

4.2

Gas Type Dependence

Pressure range

Measurement
principle

Gas type dependence

10 … 1500 mbar

diaphragm
capacitive sensor

independent of gas
type, no correction
required

1 … 10 mbar

diaphragm
capacitive sensor
and Pirani sensor

crossover range

5×10

-5

… 1 mbar

Pirani sensor

proportional to
pressure

6

)

4.3

Switching Functions SP1, SP2

The two switching functions can be set to any pressure within
the measurement range of the gauge. A solid state relay is
provided for each switching function.

6)

The pressure reading applies to dry air, O

2

, CO and N

2

. For other

gases, it has to be converted (calibration factors (

→  24).