INFICON PCG55x ATM to Medium Vacuum Gauge User Manual

Page 48

Advertising
background image

48

tina56e1-d (2014-01)

The current ATM threshold setting
• can be read / written via the diagnostic port
• is output at the measurement signal output instead of the

pressure signal, can be measured with a voltmeter, and is
displayed on the LCD display after the <SP1> or <SP2>
button is pressed

• can be read / written via the DeviceNet, Profibus and RS485

interface.

DANGER

DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing the <SP> button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the <SP> button only if you are sure that no
damages can arise from a malfunction.

Programming ATM setpoint

Programmable parameters:
(

→  [3], [4], [5], [6])

Factor of ATM
Low Trip Enable
Low Trip Point Hysteresis

High

Trip

Enable

High Trip Point Hysteresis

Setpoint

Mode


Switching characteristics of the setpoints

→  41.

Diagnostic port

→  [3].