INFICON SKY CDGxxxD (RS232C) User Manual

Page 11

Advertising
background image

tira49e1-a (2010-02) RS232C.cp

11

The "DC Output Level" variable is used for assigning a defined offset level to the
analog output signal, e.g. for setting a certain zero offset signal level.

A "DC Output Level" >0 reduces the output range of the measurement

range 0 … 10 V by the selected offset value (F.S.R. - DC output level).

The "DC Output Level" parameter (16-Bit) consists of the high and low byte.

<data_value>×a

DC Output Level =

b

×F.S.R._Mantisse×10

(F.S.R._Exp)


Parameter

1)

Description

DC Output Level DC-Output-Signal in the selected pressure unit

(

→  11, table "Parameter").

1)

Further

parameter

→  11, table "Parameter"


<data_value>

Software version =

20

e.g. <data_value> = 20

≙ V1.0

Parameter Description

<data_value> 1 byte value (8 bit), data value in decimal format.


Parameter Description

<data_value> Zero offset measurement data, consisting of "low and high

byte" (16 bit value), data value in decimal format.

a Conversion factor for pressure units other than "Torr"

Torr:

a = 1.00

mbar:

a = 1.3332

Pa:

a = 133.32

b Factor for resolution

Page

No.

1)

b Output

signal

2

32000

10.24

V

3

32000

10.24

V

4

32767

10.00

V

F.S.R._Mantisse F.S.R.

factor

according to the "Sensor type" variable, which

has to be read separately (

→ "Read command").

F.S.R._Exp F.S.R. exponent according to "Sensor type" variable, which

has to be read separately (

→ "Read command").

1)

→  3, tabel "Structure of send string", byte no. 1


DC Output Level

Software version

Parameter

Advertising