INFICON SKY CDGxxxD (RS232C) User Manual

Page 7

Advertising
background image

tira49e1-a (2010-02) RS232C.cp

7

Description Data

Comment

Read command

0x00

Read command for variable according to
address in byte No. 2

Write command

0x10

Write command for variable according to
address in byte No. 2

Special services

0x40

Direct command (write command) without
data information, e.g. Reset, Zero adjust


Enter the address of the variable to be read/written (

→ table "Variables for bytes

No. 2 and 3").

When a variable is written (receipt string) the content of byte No. 3 is written to the
variable addressed in byte No. 2 (

→ table "Variables for bytes No. 2 and 3").

When a variable is read (send string), the value of the variable addressed in byte
No. 2 is output in byte No. 6 of the send string. The content of byte No. 3 is not
relevant for read operations.

The checksum is calculated from the sum of byte No. 1 to 3.

Das Beispiel basiert auf dem Empfangsstring:

Byte

No.

0 1 2 3 4

Value 3 0 2 0 2

The instrument or controller (receiver) interprets this string as follows:

Byte No. Designation

Value Comment

0 Data

string

length

3 (constant

value)

1 Service

command

0 Read

command

2 Address

byte 2

Filter

3 Data

byte

0

4 Checksum

2

0 + 2 + 0 = 2

dec

00 02

hex

High bytes are ignored

Checksum = 02

hex

2

dec


Service command
(byte No. 1)

Address byte
(byte No. 2)

Data byte
(byte No. 3)

Checksum
(byte No. 4)

Example