INFICON SKY CDGxxxD (RS232C) User Manual

Page 5

Advertising
background image

tira49e1-a (2010-02) RS232C.cp

5

All variables in a receipt string that are addressed for reading are output on this
byte. For variable types >1 byte, each byte (e.g. low, high, or further bytes) has to
be addressed and read individually.

Read Command L-Byte

→ Read Data L-Byte

Read Command H-Byte

→ Read Data H-Byte

• After a write operation, the value of the addressed variable is output.
• After a reset (Power on) the software version is output on byte 6.


Bit No.

0 … 3

Description Comment

0x0 F.S.R.=

10

-3

→ Variable "Sensor_pressure_range"

0x1 F.S.R.=

10

-2

F.S.R._Exponent = (<value> -3)

0x2 F.S.R.=

10

-1

0x3 F.S.R.=

10

0

= 1

0x4 F.S.R.=

10

1

= 10

0x5 F.S.R.=

10

2

= 100

0x6 F.S.R.=

10

3

= 1000

0x7 F.S.R.=

10

4

= 10000

Bit No.

4 … 7

Description Comment

0x0

F.S.R._Mantissa = 1.0

→ Variable "Sensor_FSR"

0x1

F.S.R._Mantissa = 1.1

0x2

F.S.R._Mantissa = 2.0

0x3

F.S.R._Mantissa = 2.5

0x4 F.S.R._Mantissa=

5.0


The recipient (master) is synchronized by checking three bytes:

Byte No. Function

Value

Comment

0

Data string length

7

Constant value

1 Page

No.

4
3
2

Constant value for CDG025D

1)

Constant value for CDGxxxD

2)

Constant value for CDG025D

3)

8

Checksum of bytes
No. 1 … 7

0 … 255

Low byte of checksum

4)

1)

CDG025D with output signal 10.00 V (analog)

2)

CDG045D, CDG100D, CDG160D, CDG200D with output signal 10.24 V (analog)

3)

CDG025D with output signal 10.24 V (analog)

4)

Possible high bytes are ignored

Read command
(byte No. 6)

Sensor type
(byte No. 7)

Checksum and synchronization
(byte No. 8)