INFICON SKY CDGxxxD (RS232C) User Manual

Page 9

Advertising
background image

tira49e1-a (2010-02) RS232C.cp

9

(completed)

Parameter name

Data type

Description

Byte No. 2 Byte No. 3 Comment

Extended error

uint8_T / Ro

H-Byte

54

Bit 0

PT1000 fault

4)

H-Byte

Bit 1

Heater block overtemp.

4)

Bit 2

Electronic overtemp.

4)

Bit 3

Zero adjust error

Bit

4

Reserve

Bit

5

Reserve

Bit

6

Reserve

Bit

7

Reserve

Extended error

uint8_T / Ro

L-Byte

55

Bit 0

Atm. pressure out of range

L-Byte

Bit 1

Temperature out of range

Bit

2

Reserve

Bit

3

Reserve

Bit 4

Cal. mode wrong

Bit 5

Pressure underflow

Bit 6

Pressure overflow

Bit 7

Zero adjust warning

Pressure range

uint8_T / Ro

56

0

F.S.R. = E-3

(Exponent)

1

F.S.R.

=

E-2

2

F.S.R.

=

E-1

3

F.S.R. = E 0

4

F.S.R.

=

E+1

5

F.S.R.

=

E+2

6

F.S.R.

=

E+3

7

F.S.R.

=

E+4

Pressure range

uint8_T / Ro

57

0

Mantissa = 1.0

(Mantissa)

1

Mantissa = 1.1

2

Mantissa = 2.0

3

Mantissa = 2.5

4

Mantissa = 5.0

Gauge config

uint8_T / Ro

58

0

= Analog out 0 … 10.24 V

1

= Analog out 1 … 9 V

CDG type

uint8_T / Ro

59

0

= CDG025D

1

=

CDG045D

2

=

CDG100D

3

=

CDG160D

4

=

CDG200D

RW = Read / Write

Ro = Read only

1)

Factory

setting

2)

Conversion

→ Section 1.1, byte 6 "Read command"

3)

Conversion

→ Section 1.1, bytes 4 and 5 "Pressure unit"

4)

CDG045D, CDG100D, CDG160D, CDG200D only