INFICON SKY CDG100D 0.1 ... 1000 Torr / mbar User Manual

Page 11

Advertising
background image

tina52e1-d (2014-02)

11

Remote Zero Adjust


External switching contact

digital input for zero adjust-
ment (

→  25)

30 V (dc) / <5 mA (dc)

Switching functions

SP1, SP2

Setting range

0 … 99% F.S. (0 … 9.9 V)

Hysteresis 1%

F.S.

Relay contact

30 V (dc)/ ≤0.5 A (dc)
floating (NO)

closed p

≤ p

SP

(LED lit solid)

open p

≥ p

SP

(LED off)

Switching time

≤50 ms

Status relay

Relay contact

30 V (dc)/ ≤0.5 A (dc)
connected to supply com-
mon (pin 5)

closed measurement

mode

warning

open no

supply voltage

warming up
error

RS232C interface

Transmission rate
Data format


9600 baud
binary
8 data bits
one stop bit
no parity bit
no handshake
→ "Power Connection"

For further information on the RS232C interface

→  [4].

Diagnostic port

Jack connector, 2.5 mm,
3-pin

Advertising