INFICON SKY CDG100D 0.1 ... 1000 Torr / mbar User Manual
Page 26
Advertising
26
tina52e1-d (2014-02)
Operate the gauge for at least 90 minutes under constant
ambient conditions (until the signal is stable).
Briefly press the <ZERO> button with a pin (max.
ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED blinks until the adjustment (duration ≤8 s)
is completed.
Press the button briefly
max. ø1.1 mm
After zero adjustment, the gauge automatically returns
to the measurement mode.
The <STATUS> LED blinks green if
• the signal output is negative (< -20 mV) when the final pres-
sure has been attained
• the zero adjustment has failed.
4.2.2
<ZERO> Adjustment with Ramp Function
The ramp function allows to adjust the zero at a known reference
pressure within the measurement range of the gauge.
Advertising