8 supplement, 1 technical data, 1 analog input – JUMO 707060 Wtrans B - Programmable Head Transmitter with Radio Transmission Operating Manual User Manual
Page 37: Supplement, Technical data

37
8 Supplement
8.1
Technical data
8.1.1
Analog input
RTD temperature probe
Designation
Standard
Measuring range
Measuring accuracy
Pt100
DIN EN 60751
-100 to +200 °C
± 0.1 K
(TK value = 3.85 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Pt500
DIN EN 60751
-100 to +200 °C
± 0.1 K
(TK value = 3.85 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Pt1000
DIN EN 60751
-100 to +200 °C
± 0.1 K
(TK value = 3.85 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Ni 100
DIN 43760
-60 to +250 °C
±
0.2 K
(TK value = 6.18 × 10
-3
1/K)
Ni 500
DIN 43760
-60 to +150 °C
±
0.2 K
(TK value = 6.18 × 10
-3
1/K)
Ni 1000
DIN 43760
-60 to +150 °C
±
0.2 K
(TK value = 6.18 × 10
-3
1/K)
Pt100
JIS 1604
-100 to +200 °C
± 0.1 K
(TK value = 3.917 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Pt50
ST RGW 1057 19
85
-200 to +600 °C
±
0.2 K
(TK value = 3.91 × 10
-3
1/K)
Pt100
GOST 6651-
94 A.1
-100 to +200 °C
± 0.1 K
(TK value = 3.91 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Cu50
GOST 6651-
94 A.4
-50 to +200 °C
±
0.2 K
(TK value = 4.26 × 10
-3
1/K)
Cu100
GOST 6651-
94 A.4
-50 to +200 °C
±
0.2 K
(TK value = 4.26 × 10
-3
1/K)
Connection type
2-wire or 3-wire circuit
Sensor current
< 0.5 mA
Lead compensation
Not required for 3-wire circuit (admissible max. 11
Ω
per line).
Max. configurable lead-wire resistance for 2-wire circuit: 22
Ω
.