Electrical data

Technical data type 707021/..., type 707022/... and type 707025/...

Input for thermocouple

Input for resistance thermometer

Supply voltage

24V DC +10%/-15%

Power consumption

1W

Supply voltage error

 ± 0.01% per V deviation from 24V

Test voltage

to DIN 61010, Part 1

510V/50Hz, 1min

Isolation
- between input and output
- between input and mains supply
- between output and mains supply
- between input and setup plug

50V
50V
50V

no isolation between input and setup plug

Designation

Range limits

Range

Accuracy

a

Fe-Con L

DIN

43710

-200 to

+900°C

-200 to

+900°C

0.1%

above

-150°C

Fe-Con J

EN

60584

-210 to +1200°C

-200 to +1200°C

0.1%

above

-100°C

Cu-Con U

DIN

43710

-200 to

+600°C

-200 to

+600°C

0.1%

above

-100°C

Cu-Con T

EN

60584

-270 to

+400°C

-200 to

+400°C

0.1%

above

-100°C

NiCr-Ni K

EN

60584

-270 to +1372°C

-200 to +1372°C

0.1%

above

-60°C

NiCr-Con E

EN

60584

-270 to +1000°C

-200 to +1000°C

0.1%

above

-60°C

NiCrSi-NiSi N EN

60584

-270 to +1300°C

-100 to +1300°C

0.1%

above

-80°C

Pt10Rh-Pt

S

EN

60584

-50 to +1768°C

-50 to +1768°C

0.15% above 0°C

Pt13Rh-Pt

R

EN

60584

-50 to +1768°C

-50 to +1768°C

0.15% above 0°C

Pt30Rh-Pt6Rh B

EN

60584

0 —

1820°C

400 —

1820°C

0.15% above 400°C

W3Re-W25Re D

0 —

2495°C

500 —

2495°C

0.15% above 500°C

W5Re-W26Re C

0 —

2320°C

500 —

2320°C

0.15% above 500°C

Shortest span

Type L, J, U, T, K, E, N: 100°C; type S, R, B, D, C: 500°C

Range start/end

freely programmable within the limits in 0.1°C steps

Cold junction

Pt100 internal or external cold junction (adjustable from 0 to 100°C)

Cold junction accuracy

± 1°C

Sampling rate

 100msec

Special features

also programmable in °F; input isolated from output

a

The accuracy refers to the maximum range span.
For small ranges, as well as for short spans, the linearization accuracy is reduced.

Designation

Connection circuit

Range limits

Range

Accuracy

Pt 100

EN 60751

2/3-wire
2/3-wire
4-wire
4-wire

-200 to

+850°C

-100 to

+200°C

-200 to

+850°C

-100 to

+200°C

-200 to

+850°C

±0.4°C
±0.8°C
±0.4°C
±0.5°C

Pt 100

JIS

2/3-wire
2/3-wire
4-wire
4-wire

-200 to

+649°C

-100 to

+200°C

-200 to

+649°C

-100 to

+200°C

-200 to

+649°C

±0.4°C
±0.8°C
±0.4°C
±0.5°C

Pt 500

DIN

2/3-wire
2/3-wire
4-wire
4-wire

-200 to

+850°C

-100 to

+200°C

-200 to

+850°C

-100 to

+200°C

-200 to

+850°C

±0.4°C
±0.8°C
±0.4°C
±0.5°C

Pt 1000

DIN

2/3-wire
2/3-wire
4-wire
4-wire

-200 to

+850°C

-100 to

+200°C

-200 to

+850°C

-100 to

+200°C

-200 to

+850°C

±0.4°C
±0.8°C
±0.4°C
±0.5°C

Ni 100

2/3-wire
4-wire

-60 to

+180°C

-60 to

+180°C

-60 to

+180°C

±0.8°C
±0.5°C