Leica IP C

4.5 Automated unload station (optional)

Optionally available for the printer is an automat-

ed multi-level cassette unload station, where the

imprinted cassettes are collected on individually

removable and stackable trays (40) in the order

in which they were printed.
The multi-cassette unload station comes com-

plete with 10 trays, all of which can be inserted

simultaneously. Each tray holds up to 10 cas-

settes.
Installing the multi-cassette unload station:
• Unpack the automated unload station and set

it up at the designated location.

35

36

37

38

39

Fig. 18

40

36

4. Instrument Setup

Important!

Prior to installation, the printer must

be switched off and unplugged from

power supply.

The manual removal system described

in

Chapter 4.4

must not be mounted

prior to setup. The collar screws (30 in

Fig. 18

) also have to be removed.

• Place the instrument onto the unload station.

2 persons are required to do this!
• Hold the printer on both sides (right and left)

and place it so that the two rear bolts (35) fit

into the baseplate first, as shown in

Fig. 18

.

Then carefully lower the front part of the

printer unto the third bolt (36) so that the plug

connection (37) locks into place in the printer

base plate and the printer remains securely

fastened on the unload station.

• Place the stack of trays (39) onto the lifting

table (38) of the automated unload station.

See

chapter 5.2

for details on the lifting table

controls.