2 cleaning the print head, Cleaning the print head, Cleaning and maintenance – Leica Biosystems IPC User Manual

Page 56

Advertising
background image

54

Instructions for Use V 1.9 RevD – 05/2013

The print head must be cleaned

manually once a week or if the

message "15" is displayed.

3

Preparing the printer

• Open the printer lid and then

press the CLEAN and LOADED

keys simultaneously.

• The print head (72) moves

upward to a position approx.

1 cm away from the sealing lip

(

Fig. 57

).

6.2 Cleaning the print head

Fig. 57

6. Cleaning and Maintenance

69

70

72

+

Fig. 58

• Push the lever (69) upwards, then remove

the red location plate (70) with the sealing lip

(

Fig. 58

).

• Moisten one of the foam swabs supplied with

the instrument (89) with some alcohol. Be sure

not to use too much alcohol—no alcohol may

drip into the instrument.

Fig. 59

• Carefully insert the swab into the gap under

the print head (

Fig. 59

). Apply light pressure

upwards (on the print head) and move the foam

swab back and forth (approx. 10 times).

This procedure removes dried ink residues.

89

Never rotate the swab—this can dam-

age the nozzle plate of the print head.

Never use acetone or xylene! Only

use 95 % or 100 % alcohol for cleaning

purposes.