7 filling and inserting the magazines, Filling and inserting the magazines, Instrument setup – Leica Biosystems IPC User Manual

Page 32

Advertising
background image

30

Instructions for Use V 1.9 RevD – 05/2013

4. Instrument Setup

4.7 Filling and inserting the magazines

50

53

54

Depending on the type of cassette used, addi-

tional inserts must be placed into the magazine

(54) to guide the cassettes properly:
This includes: 53 – Metal insert

50 – Adhesive strip (2 mm thick)

Fig. 27

Metal strip
Insert the metal strip (53) so that the angled part

(58) touches the back panel (56) of the magazine

(54,

Fig. 29

).

51

58

Inserting the strips

56

Adhesive strip 2 mm (50)
• Remove protective foil and attach adhesive

strip in magazine so that it is centered in the

front part of the magazine and fits flush against

the bottom of the magazine foot (55,

Fig. 28

).

55

50

Fig. 28

Fig. 30

51
52

Then push the metal insert downwards until the

two plugs (51) snap into the two grooves (52) in

the magazine foot (

Fig. 30

).

58

56

53

55

Fig. 29