2 display indications, Display indications, Operation – Leica Biosystems IPC User Manual

Page 48

Advertising
background image

46

Instructions for Use V 1.9 RevD – 05/2013

5. Operation

5.2 Display indications

Display indication Magazine empty (in combination with "Mag. Empty" LED)

1 - Magazine No. 1 empty

4 - Magazine No. 4 empty

2 - Magazine No. 2 empty

5 - Magazine No. 5 empty

3 - Magazine No. 3 empty

6 - Magazine No. 6 empty

If "MANUAL FEED" has been selected in the printer driver settings, "0" will

appear in the display after the print job has been sent. The printer will wait

for an individual specimen slide to be placed in the feeding chute for printing.

Display

Status messages

00 Ink print head cleaning in progress.

11 Temperature in the flash power supply is too high.

Printer is too hot and initiates a short cool-down period.

The print job will be resumed automatically after a short period of time.

To prevent frequent job interruptions due to heat build-up, make sure

the ventilation grids of the printer are unobstructed and keep the

printer away from other heat sources.

Consider operating the printer in an air-conditioned room.

If the temperature does not drop to a value within the allowed range

within 10 minutes, "55" is displayed. Switch the instrument off and let

it cool; check ambient temperature.

12 Drying module temperature too low/too high.

If the temperature does not drop to a value within the allowed range

within 6 minutes, "43" is displayed.

13 Flash bulb has reached its maximum life.

The flash bulb has reached the end of its specified service life and

must be replaced.

If this message is ignored, the resistance of subsequent printouts

can be affected.