Trouble shooting

Systematically try several clearance
angle adjustments, until the optimum
angle is found.

-

Check if all levers are locked and
screws are tightened on the speci-
men and knife holder systems. Re-
tighten the levers and screws if nec-
essary.

-

Use a different part of the knife or
blade edge or use a new knife/blade.

-

Use a different part of the knife or
blade edge or use a new knife/blade.

-

Precool the specimen on a cold
plate.

-

Cool the specimen in iced water im-
mediately before sectioning.

-

Systematically decrease the clear-
ance angle until the optimum adjust-
ment is obtained.

-

Systematically decrease the clear-
ance angle until the optimum adjust-
ment is obtained.

-

Use knife with different profile.

-

Check if all levers are locked and
screws are tightened on the speci-
men and knife holder systems. Re-
tighten the levers and screws if nec-
essary.

-

Turn the coarse feed wheel clock-
wise to lower the specimen.

-

Insufficient knife/blade inclination;
consequently the clearance angle is
too small.

-

Insufficient clamping of specimen
and/or knife.

-

Blunt knife/blade.

-

Blunt knife/blade.

-

Specimen too warm.

-

Clearance angle too big.

-

Clearance angle too big.

-

Inappropriate knife profile.

-

Insufficient clamping of specimen
and/or knife.

-

Upper feed limit reached.

Thick/thin sections
The section thickness varies from one
section to another. In extreme cases,
sections are skipped meaning that a
section is not obtained.

Compressed sections
The sections are extremely com-
pressed, wrinkled or jammed together.

Sections exhibit scratches and chatter
marks.

No specimen feed takes place. No sec-
tion obtained.

Possible cause

Problem

Correction