Diopter settings and parfocality: 1 adjustable & 1 fixed eyepiece (continued)
9. Turn the eyelens of the eyepiece as far as it
will go in the "+" direction, without looking
into the eyepieces while doing so.
10. Close your eye on the fixed eyepiece and
look with the other eye into the into the
adjustable eyepiece.
11. Rotate the eyelens of the eyepiece slowly in
the "–" direction until the eye can see the
specimen sharply.
Continued on next page
6. Set the highest magnification level.
7. Optimize the focusing with the focusing
drive.
8. Set the lowest magnification level.
3. Place a flat specimen under the objective.
4. Set the lowest magnification level.
5. Observe the specimen through the eyepi-
eces and bring it into sharp focus with the
focusing drive.