Checking parfocality

1. Select the highest magnification level.

2. Monitor the specimen; if necessary, refocus

it slightly.

3. Change from the highest to the lowest

magnification. The sharpness should be
constant (parfocal). If this is not the case,
repeat this procedure.

Dioptric correction with two adjustable eyepieces (continued)