The field of view is shadowed

O

Adjusting the correct Interpupillary
Distance (P. 52).

The image goes out of focus.

O

Inserting the eyepieces correctly (P. 55).

O

Perform diopter correction exactly accor-
ding to the instructions (from P. 31).

The focusing drive gradually sinks on its
own or is difficult to turn.

O

Regulate the ease of movement (P. 30).

In the case of failures of electrically opera-
ted devices, always first check:

O

Is the voltage selector set correctly?

O

Is the main power switch switched on?

O

Is the power cable connected correctly?

O

Are all connecting cables attached
correctly?

O

Are the fuses intact?

Photos are blurry.

O

Focus accurately (P. 23).

O

Bring the reticule into sharp focus and
perform diopter correction exactly accor-
ding to the instructions (P. 37).

O

Insert the eyepieces correctly up to the stop
(P. 35).

O

Check that the graticules are securely in
place in the eyepiece (P. 31).

The image from the camera stays black

O

Switch the beam splitter on the photo tube
to the "Doc" setting (P. 37).

Troubleshooting