MTS DCPD Measurement System User Manual

Page 28

Advertising
background image

1. For Display Units, select "V" for Volts.

6. Assign and configure the DUC for DCPD Channel 2.

a) In the left navigation pane, go to Auxiliary Outputs.
b) Under the Input Hardware Resources tab, select the corresponding channel hardware location.

For example, 494.26 DUC S2-J1B

c) Assign the channel for the Active specimen to Aux Input 1.

1. In the Display Name field, click the plus button.

Change the Display Name to REF pd.

Change the Internal Name to REF pd.

d) Configure the settings in the General tab.

1. For Dimensions, select: "Volts".

1. For Display Units, select "V" for Volts.

7. Assign and configure the digital outputs.

a) In the navigation pane, select Digital Outputs.
b) Under the Output Hardware Resources tab, select the Dig Out 1. For example, Dig Out 1-J55 for

a 494 FlexTest controller.

1. In the Display Name field, click the plus button.

Change the Display Name to Off / On.

c) Under the Output Hardware Resources tab, select the Dig Out 2. For example, Dig Out 2-J55 for

a 494 FlexTest controller.

1. In the Display Name field, click the plus button.

Change the Display Name to Reverse.

d) Save the file.

1. Go to File > Save As.

2. Pick a file name. For example, Landmark_Frame1_DCPD.cfg.

3. Click Save.

4. If a pop-up appears, click Save Parameters.

8. Close Station Builder.

9. Configure Station Manager.

Note:

To make the changes required in this step, the Access Level must be set to Configuration. If
prompted, enter the appropriate password.

a) Open Station Manager.

28 | DCPD Measurement System